FastFinder 3.6 known issues
- When the user is using an older version of the software and the environment is updated to a newer version during the usage of the software, an error message is shown instead of a popup informing the user to logout. The user has to close the system but after restarting, FastFinder migrates successfully.
- When restarting an analysis, the analysis name is reset to the original file name, and not the changed analysis name (if appropriate).
- When an Assay Plugin defines the same nametag for different sample types, changing these nametags in FastFinder is not possible. A new assay plugin with the correct nametags must be imported to correct the nametags.
- Warning dialog about unsaved settings does not appear when selecting or deselecting the "plate" column in the results overview settings.
- Edge cases of the Ctrl-shift behavior in the plate view do not work as expected.
- Keyboard navigation in the to be reviewed tab only works after clicking twice on a sample.
- The confirm checkbox in the QC overview is not automatically checked when the second reviewer changes the action for the QC violation.
- Sorting on Result in the details tab does not work correctly. This can easily be mitigated by using one or multiple other filters.
- After removing a sample that was assigned by a .flow file from the plate setup, the tooltip of that well is empty.
- Multiplate assay template loading does not work in a certain edge case
- Order of sample graphs in the pdf export for multimix assays differs from table overview.
- When selecting wells in plate view, PC/NC/reference curves are deselected
- In the CSV export the "Experiment creation date (UTC)" header is not in UTC
- In contrast to what is described in the IFU, BioRad® CFX 384 is not supported.
- For a correct application of crosstalk correction for Agilent’s AriaMx and AriaDx files, software version 1.5 must be used
- Lot expiry date is not visualized correctly (Date is not converted from local time to UTC)
FastFinder 3.5 known issues
Risk analysis of below-known anomalies revealed that there is only a low risk associated with these anomalies and that the system could be released without resolving them. These anomalies will be fixed in future validated releases of the FastFinder Platform software.
- When the user is using an older version of the software and the environment is updated to a newer version during the usage of the software, an error message is shown instead of a popup informing the user to logout. The user has to close the system but after restarting, FastFinder migrates successfully.
- When restarting an analysis, the analysis name is reset to the original file name, and not the changed analysis name (if appropriate).
- When an Assay Plugin defines the same nametag for different sample types, changing these nametags in FastFinder is not possible. A new assay plugin with the correct nametags must be imported to correct the nametags.
- Warning dialog about unsaved settings does not appear when selecting or deselecting the "plate" column in the results overview settings.
- Edge cases of the Ctrl-shift behavior in the plate view do not work as expected.
- Keyboard navigation in the to be reviewed tab only works after clicking twice on a sample.
- The confirm checkbox in the QC overview is not automatically checked when the second reviewer changes the action for the QC violation.
- Sorting on Result in the details tab does not work correctly. This can easily be mitigated by using one or multiple other filters.
- After removing a sample that was assigned by a .flow file from the plate setup, the tooltip of that well is empty.
- Multiplate assay template loading does not work in a certain edge case
- Order of sample graphs in the pdf export for multimix assays differs from table overview.
- In the CSV export the "Experiment creation date (UTC)" header is not in UTC
- Lot expiry date is not visualized correctly (Date is not converted from local time to UTC)